Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med PM-10: (1 year) CAL VER /5520 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 130 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Astro-Med PM-10 Calibration Procedure TP# 3163000X Rev D. # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N P J F W 1.003 ASK+ R X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD SET-UP PROCEDURE 1.008 MESS Slew the 5520A for full scale deflection. 1.008 MESS 1.009 DISP Insure that the PM-10 is not installed in a 1.009 DISP channel directly following a PM-08. 1.009 DISP 1.009 DISP Press Zero Defaults. 1.010 DISP Note the position of DSW1 dip switches to determine 1.010 DISP the type of thermocouple the PM-10 is configured for. 1.010 DISP 1.010 DISP K = OFF ON OFF OFF 1.010 DISP J = OFF ON OFF ON 1.010 DISP E = OFF ON ON OFF 1.010 DISP T = OFF OFF OFF ON 1.011 MEM2 Enter thermocouple type (K, J, E, or T). 1.012 JMPL TYPE_K ZCMPI(MEM2, "K") 1.013 JMPL TYPE_J ZCMPI(MEM2, "J") 1.014 JMPL TYPE_E ZCMPI(MEM2, "E") 1.015 JMPL TYPE_T ZCMPI(MEM2, "T") 1.016 JMP 1.010 1.017 LABEL TYPE_K 1.018 HEAD {TYPE K THERMOCOUPLE} 1.019 DISP Connect the 5520A TC Terminals to the PM-10 input using 1.019 DISP TYPE K thermocouple wire and connectors. 1.020 HEAD TYPE K THERMOCOUPLE: {50deg Range} 1.021 5520 50 50degC +1.25% _K TC 2W 2.001 HEAD TYPE K THERMOCOUPLE: {100deg Range} 2.002 5520 100 100degC -0.63% _K TC 2W 3.001 HEAD TYPE K THERMOCOUPLE: {200deg Range} 3.002 5520 200 200degC +1% _K TC 2W 4.001 HEAD TYPE K THERMOCOUPLE: {500deg Range} 4.002 5520 500 500degC -0.1% _K TC 2W 5.001 HEAD TYPE K THERMOCOUPLE: {1000deg Range} 5.002 5520 1000 1000degC -0.2% _K TC 2W 6.001 JMP 18.001 6.002 LABEL TYPE_J 6.003 HEAD {TYPE J THERMOCOUPLE} 6.004 DISP Connect the 5520A TC Terminals to the PM-10 input using 6.004 DISP TYPE J thermocouple wire and connectors. 6.005 HEAD TYPE J THERMOCOUPLE: {50deg Range} 6.006 5520 50 50degC -6% _J TC 2W 7.001 HEAD TYPE J THERMOCOUPLE: {100deg Range} 7.002 5520 100 104degC -4% _J TC 2W 8.001 HEAD TYPE J THERMOCOUPLE: {200deg Range} 8.002 5520 200 202degC -1% _J TC 2W # Changed tolerance for 500degC J type from 0% to .001U so that procedure # would pass if 500degC was entered but still fail if anything other than # 500degC is entered. 9.001 HEAD TYPE J THERMOCOUPLE: {500deg Range} 9.002 5520 500 500degC 0.001U _J TC 2W 10.001 HEAD TYPE J THERMOCOUPLE: {1000deg Range} 10.002 5520 1000 950degC +4% _J TC 2W 11.001 JMP 18.001 11.002 LABEL TYPE_E 11.003 HEAD {TYPE E THERMOCOUPLE} 11.004 DISP Connect the 5520A TC Terminals to the PM-10 input using 11.004 DISP TYPE E thermocouple wire and connectors. 11.005 HEAD TYPE E THERMOCOUPLE: {50deg Range} 11.006 5520 50 53degC -20% _E TC 2W 12.001 HEAD TYPE E THERMOCOUPLE: {100deg Range} 12.002 5520 100 105degC -17% _E TC 2W 13.001 HEAD TYPE E THERMOCOUPLE: {200deg Range} 13.002 5520 200 222degC -11% _E TC 2W 14.001 HEAD TYPE E THERMOCOUPLE: {500deg Range} 14.002 5520 500 459degC -5% _E TC 2W 15.001 JMP 18.001 15.002 LABEL TYPE_T 15.003 HEAD {TYPE T THERMOCOUPLE} 15.004 DISP Connect the 5520A TC Terminals to the PM-10 input using 15.004 DISP TYPE T thermocouple wire and connectors. 15.005 HEAD TYPE T THERMOCOUPLE: {50deg Range} 15.006 5520 50 55degC -30% _T TC 2W 16.001 HEAD TYPE T THERMOCOUPLE: {100deg Range} 16.002 5520 100 105degC -25% _T TC 2W 17.001 HEAD TYPE T THERMOCOUPLE: {200deg Range} 17.002 5520 200 200degC -15% _T TC 2W 18.001 END